DEVELOPMENT OF A DYNAMIC BALLOON VOLUME SENSOR SYSTEM FOR USE IN PULSATING BALLOON CATHETERS WITH CHANGING HELIUM CONCENTRATIONS

Nolan, Timothy David Campbell (2007) DEVELOPMENT OF A DYNAMIC BALLOON VOLUME SENSOR SYSTEM FOR USE IN PULSATING BALLOON CATHETERS WITH CHANGING HELIUM CONCENTRATIONS. Master's Thesis, University of Pittsburgh. (Unpublished)

[img]
Preview
PDF
Primary Text

Download (5MB) | Preview

Abstract

A dynamic balloon volume sensor system (DBVSS) was designed for use with the intra-aortic balloon pump (IABP), a therapeutic device to assist heart recovery after cardiac dysfunction or cardiac trauma, and the Pittsburgh respiratory support catheter (RSC), an internally deployed gas exchange device which augments lung function. The DBVSS was designed to detect the degree of inflation of the balloons incorporated into each device as they pulse within a patient. Both devices require full inflation for optimal performance, and both will under-inflate during normal operation. The sensor system requirements were to measure volumes within 10% of the actual across the range of expected pulsation frequencies as well as in changing concentrations of helium.The DBVSS employed a hot wire anemometer to detect the flow entering the balloon, combined with a computer algorithm to integrate the flow to find volume. The system compensated for the flow reading changes resulting from changing helium concentration by measuring gas properties during zero gas flow between pulsations, and used this data to correct the flow profile at each helium concentration. The volume from the DBVSS was compared to the volume standard as measured by water displacement in a plethysmograph.The system was able to accurately measure delivered balloon volume under changing gas composition as well as detected volume loss from the balloon across helium concentrations. The DBVSS measured the volume within 10% across these tests, as well as under compression of the balloon, high resistance in the driveline and across frequencies up to 480 beats per minute. The DBVSS was proved to be within the design requirements for helium concentration and inflation methods for both the devices considered.


Share

Citation/Export:
Social Networking:
Share |

Details

Item Type: University of Pittsburgh ETD
Status: Unpublished
Creators/Authors:
CreatorsEmailPitt UsernameORCID
Nolan, Timothy David Campbelltnolan@smail.umaryland.edu
ETD Committee:
TitleMemberEmail AddressPitt UsernameORCID
Committee ChairFederspiel, Williamfederspielwj@upmc.eduWFEDERSP
Committee MemberStetten, Georgegeorge@stetten.com
Committee MemberAntaki, Jamesantaki@andrew.cmu.edu
Date: 31 January 2007
Date Type: Completion
Defense Date: 14 November 2003
Approval Date: 31 January 2007
Submission Date: 7 December 2004
Access Restriction: No restriction; Release the ETD for access worldwide immediately.
Institution: University of Pittsburgh
Schools and Programs: Swanson School of Engineering > Bioengineering
Degree: MSBeng - Master of Science in Bioengineering
Thesis Type: Master's Thesis
Refereed: Yes
Uncontrolled Keywords: helium flow; hollow fiber oxygenator; spirography
Other ID: http://etd.library.pitt.edu/ETD/available/etd-12072004-213820/, etd-12072004-213820
Date Deposited: 10 Nov 2011 20:08
Last Modified: 19 Dec 2016 14:38
URI: http://d-scholarship.pitt.edu/id/eprint/10127

Metrics

Monthly Views for the past 3 years

Plum Analytics


Actions (login required)

View Item View Item